©American
Society for Veterinary Clinical Pathology |
|
Principles
of Quality Assurance and Standards
for
Veterinary Clinical Chemistry
I.
Preanalytical Factors Important in Clinical Chemistry
II.
Analytical Factors Important in Clinical Chemistry
III.
Postanalytical Factors Important in Clinical Chemistry
I.
Preanalytical Factors Important in Clinical Chemistry [Return
to Top]
A.
Specimen Collection, Handling, and Transport to the Laboratory
Samples should
be appropriately collected, handled and transported to the laboratory
in a timely manner, dependent on the type of specimen and its stability.
For any assay performed in the laboratory, information concerning sample
requirements, proper collection, handling, and delivery or shipping procedures
should be available to clients in a laboratory services manual, special
information sheets, journal or newsletter articles, other written materials,
or by personal or telephone conversation.
B.
Specimen Identification
Specimens
should be identified with pertinent information as determined by the laboratory
(such as owner, species, animal signalment, name of clinic or doctor,
address, telephone and fax numbers, e-mail address, location from which
the specimen was collected, etc.) on the submission container and submission
form.
C.
Test Identification
The requested
test(s) should be clearly stated on the submission form.
D.
Specimen Accessioning
The specimen
should be correctly entered into the laboratory system. Test request entry,
delivery of the specimen to the correct location, and specimen aliquoting
(if necessary) or sharing between laboratories or departments (ie, pharmacology,
endocrinology, and clinical chemistry) should be coordinated.
E.
Client Communication and Education
Communication
between laboratory personnel and clients should be timely and courteous
regarding pre-analytical factors influencing laboratory test results (eg,
incomplete submission forms, inappropriate sample or sample handling or
poor sample quality). Clients should be informed of the expected time
for receipt of preliminary and final reports.
F.
Personnel Safety
Personal
protective equipment should be appropriate for handling specimens and
equipment used for clinical chemistry. Safety procedures and disposal
of all samples and supplies should be appropriate for the type of specimen.
Personnel should receive safety and biohazard training and information
about exposure to potentially hazardous chemicals or infectious agents.
All training should be documented.
G.
Laboratory Environment
The laboratory
space should be clean, well lit, and organized to ensure proper achievement
of the above goals.
H.
Personnel Requirements
Laboratory
personnel should have training in specimen handling and sample preparation.
Documentation of training, continuing education and periodic proficiency
assessment should be at the discretion of the laboratory director.
Return
to Top
II.
Analytical Factors Important in Clinical Chemistry [Return
to Top]
A.
Monitoring
1. Internal
monitoring should include the following:
(a) Quality
of water (as specified by instrumentation and assays).
(b) Stability
of electrical power (as specified by instrumentation).
(c) Temperatures
of water bath, refrigerator, and freezer (recommended at least monthly).
(d) Regular
calibration of analytical balances and pipettes (recommended annually).
(e) Maintenance
of up-to-date procedure manuals with clearly stated dates when procedures
are first implemented and when any changes are made and implemented.
(f) Maintenance
of adequate inventory, with proper storage and handling.
(g) Maintenance
of a log of changes in any procedures, problems or other factors affecting
methods, as well as actions that resolved the problem. All entries
should be clearly dated and signed by laboratory personnel.
2. External
monitoring should include participation in an external proficiency program
(a) All
participating laboratories should analyze the same materials.
(b) Results
should be tabulated regularly (monthly, quarterly) and distributed
to participants with statistical summaries and comparison of participating
laboratories with mean indices expressing the closeness of individual
laboratory results to the group mean.
(c) Means
should be calculated and analyzed based on identification of the method
(same methods compared).
(d) Each
laboratory should carefully assess the validity of their reported
performance and consider any changes indicated by the proficiency
program.
B.
Method Validation (Return
to Top)
Method validation
should be performed before a test procedure is placed into routine use.
Validation may be accomplished by thoroughly testing reference materials
or by comparison of results of tests performed by an alternative method.
For each method, the laboratory should verify the manufacturer’s
claims and any adjustments before initiating patient testing.
Method validation
should provide evidence of the following:
1. Accuracy
– Perform either (a) or (b)
(a) Run
known value substance and compare results to expected value.
(b)
Perform split sample patient comparison between existing method of
known accuracy and new method .
2. Precision
– Perform either (a) or (b)
(a)
Run 10 replicates of 2 levels of quality control (QC) samples.
(b) Gather
21 results; 7 results in each of 3 separate runs (better estimate
of day-to-day precision, as well as within-run precision).
With
results from (a) or (b) determine mean, standard deviation (SD) and
coefficient of variation (CV). Determine whether within-run SD is
acceptable.
3. Sensitivity
– Perform (a), (b) or (c)
(a) Assess
manufacturer’s claims.
(b) Use
concentration of low calibrator or another sample or fluid with low
levels of analyte.
(c) Run
a series of dilutions and assess acceptability of performance.
4. Specificity
– Perform (a) or (b)
(a) Use
published list of interfering substances, check with manufacturer.
(b) Assess
known or suspected interfering substances by spiking specimens or
use patient material with known conditions.
5. Linear
reportable range
(a)
Establish upper and lower limits for reporting patient values based
on calibration materials.
(b) For
the lower limit, there should be confirmation of the discriminatory
ability of the test.
(c) The
highest calibration point is the maximum upper limit and the lowest
calibration point or zero should be the minimum lower limit for reporting
patient results.
6. Linearity
– Perform either (a) or (b)
(a) Determine
by analyzing multiple dilutions of either a high calibrator, control
or patient samples with increased levels of analyte.
(b) Analyze
calibrators of variable, known concentrations.
(c) Linearity
should be established at the time of validation and whenever new or
altered reagents are used.
7. Reference
intervals
(a) The
laboratory should establish or validate existing reference intervals
for each method and species before reporting results.
(b) Parallel
tests should be run to confirm reference intervals for controls when
changing reagents or QC lot number.
C.
Instrumentation (Return
to Top)
1. Instrument
performance
The equipment
and instrument used must be capable of providing test results within
the laboratory’s stated performance characteristics. These include:
(a)
Detection limits
(b) Precision
(c) Accuracy
(d) Specificity
(e) Sensitivity
(f) Freedom
from interferences and related test variables (refer to previous section
on method validation)
(g) Additional
points to consider:
- Instruments
with adjustable setting for different substances and/or species
should be carefully checked for compliance
- Compare
and make adjustments for performance characteristics as defined
by the laboratory and the manufacturer
- Make
certain species differences are accommodated
2. Function
checks
(a) Appropriate
function checks should be made on all instruments. These are critical
operating characteristics of an instrument, ie, stray light, zeroing,
electrical levels, optical alignment, background checks, etc.
(b) Laboratory
personnel should recheck and/or calibrate each instrument daily or
once per shift, prior to patient testing, to ensure that it is functioning
correctly and is properly calibrated. This includes daily QC.
3. Calibration
(a) Instruments
should be calibrated every 6 months or more frequently if indicated
by:
- Manufacturer’s
recommendation
- After
major service
- QC
outside limits or troubleshooting indicates need
- Laboratory
determination that volume, equipment performance or reagent stability
indicate a need for more frequent calibration
(b) After
calibration, controls should be run
4. Laboratory
personnel knowledge of equipment and its use, including, but not limited
to:
(a) Linearity
differences from possible manufacturer’s range (human) to animal
(b) Effects
of hemolysis, lipemia, icterus, caretenoid pigments (especially large
animals), and different anticoagulants on each assay
(c)
Reportable ranges
(d) Species-specific
ranges and reference intervals
(e) Expected
abnormal ranges
(f) Common
problems encountered with veterinary samples
(g) Regular
instrument maintenance schedule
(h) Replacement
of inadequate or faulty equipment
(i) Problem-solving
procedures, troubleshooting
D.
Quality Control (Return
to Top)
1. For
each run, at lest 2 controls should be assayed. Use of "high"
and "low" abnormal controls is recommended.
2. Maximum
length of a run is 24 hours. If the instrument manufacturer requires
more frequent controls, observe the recommended frequency (ie, some
blood gas instruments).
3. Verify
that the instrument is stable over the "run time". During
a validation check, controls are assayed more frequently to establish
run time.
4. Establish
QC frequency; consider the following:
(a) Test
volume (number performed each run or day) and frequency
(b) Technique
dependence of the method
(c) Analyte
or reagent stability
(d) Frequency
of QC failures
(e) Training
and experience of personnel
(f) Cost
of QC (increasing frequency adds to cost-per-test)
5. Quality
control parameters
(a) Mean,
SD and CV should be calculated (minimum n = 20).
(b) Controls
should be assayed in the same manner as patient specimens .
(c) A
mechanism should be in place to determine whether testing personnel
follow policies and procedures correctly.
(d) Use
of Westgard multirule procedures or other rules based on QC validation
is recommended.
(e) Policies
and procedures should be written and available in a laboratory Standard
Operating Procedures (SOP) manual to ensure accurate and reliable
test results .
(f) An
SOP manual should have clearly marked and dated entries of current
procedures (manufacturer package inserts are sufficient as long as
verified) and when any changes are made and implemented.
(g) QC
records should be reviewed frequently to ensure that when QC values
fail to meet the criteria for acceptability, suitable action is taken.
(h) Control
products should be purchased commercially, if possible. If using calibrators
as controls, use a different lot as QC material. If patient pooled
samples are used, establish the mean value of all analytes (minimum
n = 10 to establish a mean).
(i) Monitor
results of clinical specimens for various sources of error by use
of parameters such as anion gap, comparison of test results with previous
submissions from same patient (delta checks), and investigation of
markedly abnormal results (limit checks).
E.
Procedures Manual (Return
to Top)
1. All
procedures currently in use should be included. Protocols may be organized
in manuals and/or stored in computers, and be written form. They should
contain such information as:
(a) Patient
preparation
(b) Specimen
collection, processing and handling
(c) Criteria
for rejection of specimens
(d) Limitations
of and things that interfere with the method in use
(e) Step-by-step
procedures
(f) Reagent
preparation or manufacturer
(g) Reference
interval
(h) Reportable
range
(i) Literature
references
(j) Reagent
labeling: content, storage requirements, expiration
(k) Laboratory-specific
information, such as:
- Identification
of instrument used
- Result
reporting method
- Actions
to take when system is down
- Criteria
for specimen referrals to outside laboratories ("send outs")
- Quality
control procedures
- Documentation
of critical values
- Clearly
stated and dated entries of procedure implementation or change
F.
Comparison of Test Results (Return
to Top)
If the laboratory
performs the same test by more than 1 method or at more than 1 test site,
or the test is sometimes also sent to a referral laboratory, comparisons
should be run at least twice annually to define the relationships between
methods and sites. Comparison of different test methods for the same analyte
within the laboratory or between laboratories (if samples are tested in-house
and at a referral laboratory) is recommended. This should be done every
6 months or at a frequency determined by the laboratory manager. The following
steps should be included:
1. Perform
a 20-sample or greater comparison using specimens covering the analytical
range.
(a) Group
data in an x-y comparison plot
(b) Calculate
slope and intercept by a least squares method
2. Laboratory
director or qualified personnel should define acceptable performance
limits .
3. If individual
test results performed on the same patient or material do not correlate
with each other (ie, BUN/creatinine, electrolyte balance), the cause
should be investigated and corrective action taken.
Postanalytical Factors Important in Clinical
Chemistry [Return
to Top]
A.
Computer Entry of Data
Reports should
be accurate whether created manually or electronically, and in a standard
format as established by the laboratory. Established laboratory standards
for uniform reporting should be met.
B.
Report Generation
Reports should
be in a format that is readable and easily understood, with appropriate
references or explanations as needed. They should be generated in a timely
manner relative to preanalytical and analytical components.
C.
Report Delivery
Report delivery
should be timely, to the correct client, and in a manner agreed upon by
the client and the laboratory.
D.
Specimen Storage
Specimens
should be stored under appropriate conditions for a predetermined time
period, as determined by specimen stability, laboratory policy and/or
certification/accreditation requirements.
E.
Specimen Disposal
Laboratories
should appropriately dispose of biohazardous and non-biohazardous materials
and specimens, including timely emptying of all containers and trash bins.
F.
Personnel Safety
Conditions
should be appropriate for computer entry, transcription, handling of specimens,
specimen disposal and all other postanalytical tasks.
G.
Laboratory Environment
Laboratory
environment should meet standard requirements necessary for safe, rapid,
efficient and effective performance.
H.
Personnel Requirements
Personnel
should meet training requirements as indicated for specific areas of the
laboratory.
Return
to Top
|